Measurements of noise produced by incubators and their associated live support equipment (oxygen central network) has been performed. In 33.3% of the incubators the noise level was above the standard specification (60 dB (A)) according to International Standards. Without live support equipment, the mean noise level was 58.9 dB (A). Using live support equipment with oxygen flow of 3,5 and 81/min, the mean noise level has increased to 60.5dB (A) (P=0.53), 60.9 dB (A) (P=0.39) and 67.2 dB (A) (P=0.017) respectively, showing a remarkable influence of the central gas distribution on the incubator internal noise. In 82.5% of the incubators using a plastic box over the infant's head an injecting moist oxigen/compressed air the noise level was above the standard specification; when oxygen proportion was increased the internal incubator noise level increased as well showing a statistically significant relationship between them. Measurements of impulse noise produced by incubators lateral access door handling has shown a 107.0 dB (F).
Informação da revista
Vol. 70. Núm. 03.
Páginas 157-162 (maio - junho 1994)
Vol. 70. Núm. 03.
Páginas 157-162 (maio - junho 1994)
DOI: 10.2223/JPED.771
Acesso de texto completo
Avaliação de nível de ruído em incubadora
Visitas
1189
a Médica; Mestranda em Engenharia Biomédica, Curso de Pós-Graduação em Engenharia e Informática, CPGEI, Centro Federal de Educação Tecnológica do Paraná, CEFET-PR.
b Engenheiro Eletrônico; Laboratório de Desenvolvimento, Instituto de Tecnologia do Paraná, TECPAR; Mestrando em Engenharia Biomédica, Curso de Pós-Graduação em Engenharia e Informática, CPGEI, Centro Federal de Educação Tecnológica do Paraná, CEFET-PR.
c Doutor em Engenharia Biomédica; Professor Adjunto do Curso de Pós-Graduação em Engenharia e Informática, CPGEI, Centro Federal de Educação Tecnológica do Paraná, CEFET-PR.
Este item recebeu
Informação do artigo
Abstract
O texto completo está disponível em PDF