To investigate the causes and degree of interobserver variability in esophageal pH monitoring for the diagnosis of gastroesophageal reflux.
MethodsThis retrospective study included all children (n = 72) who underwent pH monitoring during 1 year at Maxima Medical Centre in Veldhoven, the Netherlands.
ResultsAn interobserver variability of 18% was found. Variability was caused by differences in opinion about the duration of registration, doubts about probe position, artifacts and drift of baseline pH.
ConclusionsMost of these problems can be eliminated by posttest calibration and assessment of the pH electrode position. However, a clear definition of monitoring artifacts is lacking. This study shows that mutual agreement in the interpretation of pH studies was fair (kappa coefficient of 0.70).
O monitoramento do pH esofágico é considerado o método mais confiável para diagnosticar o refluxo gastroesofágico. No entanto, a interpretação do mesmo estudo de pH pode diferir entre observadores. Neste estudo, investigamos as causas e o grau de variabilidade entre observadores.
MétodosEste estudo retrospectivo incluiu todas as crianças (n = 72) que realizaram monitoramento de pH durante 1 ano no Maxima Medical Centre, em Veldhoven, Holanda.
ResultadosFoi encontrada uma variabilidade de 18% entre observadores. A variabilidade foi causada por diferenças de opinião sobre a duração do registro, dúvidas sobre a posição da sonda, artefatos e flutuação do pH de base.
ConclusõesA maioria desses problemas pode ser eliminada por avaliação da posição do eletrodo de pH e calibração pós-teste. No entanto, ainda falta uma definição clara dos artefatos de monitoramento. Este estudo mostra que a concordância mútua na interpretação dos estudos de pH foi regular (coeficiente kappa de 0,70).